Towards a one step geometric calibration of an optical coherence tomography
- verfasst von
- Jesús Díaz Díaz, Maik Rahlves, Omid Majdani, Eduard Reithmeier, Tobias Ortmaier
- Abstract
This article presents a one step geometric calibration for an optical coherence tomography (OCT) which forms part of a medical navigation system. The 3D landmark-based geometric calibration with a self-produced 3D reference structure is based on the identification of a parameterized grey-box OCT model. We show in experimental results by comparing common measurement errors in the field of medical surgery before and after calibration, that the proposed methodology reduces systematic errors by more than one order of magnitude. Due to its simplicity, the calibration can be carried out directly before a surgical intervention enhancing the OCT accuracy.
- Organisationseinheit(en)
-
Hannoversches Zentrum für Optische Technologien (HOT)
Institut für Mechatronische Systeme
- Externe Organisation(en)
-
Medizinische Hochschule Hannover (MHH)
Exzellenzcluster Hearing4all
Vanderbilt University
Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU Erlangen-Nürnberg)
- Typ
- Aufsatz in Konferenzband
- Publikationsdatum
- 2012
- Publikationsstatus
- Veröffentlicht
- Peer-reviewed
- Ja
- ASJC Scopus Sachgebiete
- Elektronische, optische und magnetische Materialien, Physik der kondensierten Materie, Angewandte Informatik, Angewandte Mathematik, Elektrotechnik und Elektronik
- Elektronische Version(en)
-
https://doi.org/10.1117/12.999276 (Zugang:
Unbekannt)