A one step vs. a multi step geometric calibration of an optical coherence tomography

verfasst von
Jesús Díaz Díaz, Maik Rahlves, Omid Majdani, Eduard Reithmeier, Tobias Ortmaier
Abstract

This contribution compares two approaches for the geometric calibration of an optical coherence tomography (OCT) which forms part of a medical navigation system. For this purpose, a one step and a multi step calibration is performed with a self-produced 3D reference structure and a high-accurate 6 degrees of freedom (DoF) parallel robot, respectively. These 3D landmark-based geometric calibrations are based on the identification of a parameterized grey-box OCT model. We show in experimental results that both methods reduce systematic errors by more than one order of magnitude.

Organisationseinheit(en)
Hannoversches Zentrum für Optische Technologien (HOT)
Institut für Mess- und Regelungstechnik
Institut für Mechatronische Systeme
Externe Organisation(en)
Medizinische Hochschule Hannover (MHH)
Typ
Aufsatz in Konferenzband
Publikationsdatum
2013
Publikationsstatus
Veröffentlicht
Peer-reviewed
Ja
ASJC Scopus Sachgebiete
Elektronische, optische und magnetische Materialien, Biomaterialien, Atom- und Molekularphysik sowie Optik, Radiologie, Nuklearmedizin und Bildgebung
Elektronische Version(en)
https://doi.org/10.1117/12.2005264 (Zugang: Unbekannt)