A one step vs. a multi step geometric calibration of an optical coherence tomography
- verfasst von
- Jesús Díaz Díaz, Maik Rahlves, Omid Majdani, Eduard Reithmeier, Tobias Ortmaier
- Abstract
This contribution compares two approaches for the geometric calibration of an optical coherence tomography (OCT) which forms part of a medical navigation system. For this purpose, a one step and a multi step calibration is performed with a self-produced 3D reference structure and a high-accurate 6 degrees of freedom (DoF) parallel robot, respectively. These 3D landmark-based geometric calibrations are based on the identification of a parameterized grey-box OCT model. We show in experimental results that both methods reduce systematic errors by more than one order of magnitude.
- Organisationseinheit(en)
-
Hannoversches Zentrum für Optische Technologien (HOT)
Institut für Mess- und Regelungstechnik
Institut für Mechatronische Systeme
- Externe Organisation(en)
-
Medizinische Hochschule Hannover (MHH)
- Typ
- Aufsatz in Konferenzband
- Publikationsdatum
- 2013
- Publikationsstatus
- Veröffentlicht
- Peer-reviewed
- Ja
- ASJC Scopus Sachgebiete
- Elektronische, optische und magnetische Materialien, Biomaterialien, Atom- und Molekularphysik sowie Optik, Radiologie, Nuklearmedizin und Bildgebung
- Elektronische Version(en)
-
https://doi.org/10.1117/12.2005264 (Zugang:
Unbekannt)